A volumetric meter for monitoring of low gas flow rate from laboratory-scale biogas reactors

Author

Summary, in English

A reliable and inexpensive device for measuring gas with low flow rates is described. The device works based on the principle of liquid displacement and is able to measure flows from 1 to 950 ml(.)h(-1). It also has improved counting logic and data processing capacity to avoid miscounting and provide a friendly interface for data display. (C) 2003 Elsevier B.V. All rights reserved.