Development of biosensors based on hexacyanoferrates.

Author

Summary, in English

Ferric and copper hexacyanoferrates (PB and CuHCF, respectively) were electrodeposited on glassy carbon electrodes providing a suitable catalytic surface for the amperometric detection of hydrogen peroxide. Additionally glucose oxidase was immobilized on top of these electrodes to form glucose biosensors. The biosensors were made by casting glucose oxidase-Nafion layers onto the surface of the modified electrodes. The operational stability of the films and the biosensors were evaluated by injecting a standard solution (5 muM H(2)O(2) for PB, 5 mM H(2)O(2) for CuHCF and 2.5 mM glucose for both) over 5-10 h in a flow-injection system with the electrodes polarized at -50 (PB) and -200 mV (CuHCF) versus Ag/AgCl, respectively. The glucose biosensors demonstrated suitability for glucose determination: 0.0-2.5 mM (R(2)=0.9977) for PB and 0.0-10 mM (R(2)=0.9927) for CuHCF, respectively. The visualization of the redox catalyst modifiers (PB and CuHCF films) was presented by scanning electron micrographs.

Publishing year

2000

Language

English

Pages

791-799

Publication/Series

Talanta

Volume

52

Issue

5

Document type

Journal article

Publisher

Elsevier

Topic

  • Medical Engineering
  • Analytical Chemistry

Status

Published

ISBN/ISSN/Other

  • ISSN: 1873-3573