Flow-through picoliter dispenser: A new approach for solvent elimination in FT-IR spectroscopy
Author
Summary, in English
A new interface for FT-IR analysis of liquid samples on the basis of solvent elimination is presented. The approach is based on a piezoactuated flow-through microdispenser, a device built of two microstructured silicon wafers designed for micro-liquid handling. It could be verified during preliminary studies using a sequential injection (SI) system for automated liquid handling that the flow-through microdispenser as a possible interface for flow system-FT-IR analysis has the capability of meeting the demands of hyphenated miniaturized liquid handling systems (e.g., mu-HPLC, microhigh performance liquid chromatography), as it successfully provides highly stable, reliable and reproducible operating conditions for liquid handling in the picoliter range. Moreover, an increase in sensitivity for FT-IR measurements could be achieved, lowering the mass detection limit of sugars (such as the investigated sucrose) to 53 picograms. As is demonstrated on the example of an HPLC separation of a mixture of glucose and fructose, interfacing LC systems to FT-IR using a piezoactuated flow-through microdispenser is a feasible and promising approach.
Department/s
Publishing year
2002
Language
English
Pages
902-908
Publication/Series
Applied Spectroscopy
Volume
56
Issue
7
Document type
Journal article
Publisher
Society for Applied Spectroscopy
Topic
- Medical Engineering
Keywords
- micro-liquid handling
- FT-IR
- infrared spectroscopy
- Fourier transform
- flow-through microdispenser
- solvent elimination
Status
Published
ISBN/ISSN/Other
- ISSN: 1943-3530