A sub-picosecond accumulating streak camera for x-rays

Author

Summary, in English

An x-ray streak camera system uses a laser triggered photo-conductive switch to synchronize the high voltage ramp applied to its sweep plates to the photo-excitation of a sample. This technique allows the jitter. between successive sweeps to be kept below 100 fs compared to over 1 ps:using classical methods. By accumulating a stable pulsed x-ray signal over many sweeps, a very high dynamic range can be achieved whilst maintaining a sub-picosecond time resolution. The ultimate time resolution is limited by the dispersion of electrons in the streak tube. Whilst triggering with a laser at 900 Hz we have achieved a time resolution (FWHM) of 640 fs at a detection wavelength of 267 nm for a 60 s accumulation period, corresponding to over 50 000 sweeps.

Department/s

Publishing year

2001

Language

English

Pages

1858-1864

Publication/Series

Measurement Science & Technology

Volume

12

Issue

11

Document type

Journal article

Publisher

IOP Publishing

Topic

  • Atom and Molecular Physics and Optics

Status

Published

ISBN/ISSN/Other

  • ISSN: 0957-0233