A sub-picosecond accumulating streak camera for x-rays
Author
Summary, in English
An x-ray streak camera system uses a laser triggered photo-conductive switch to synchronize the high voltage ramp applied to its sweep plates to the photo-excitation of a sample. This technique allows the jitter. between successive sweeps to be kept below 100 fs compared to over 1 ps:using classical methods. By accumulating a stable pulsed x-ray signal over many sweeps, a very high dynamic range can be achieved whilst maintaining a sub-picosecond time resolution. The ultimate time resolution is limited by the dispersion of electrons in the streak tube. Whilst triggering with a laser at 900 Hz we have achieved a time resolution (FWHM) of 640 fs at a detection wavelength of 267 nm for a 60 s accumulation period, corresponding to over 50 000 sweeps.
Department/s
Publishing year
2001
Language
English
Pages
1858-1864
Publication/Series
Measurement Science & Technology
Volume
12
Issue
11
Full text
- Available as PDF - 150 kB
- Download statistics
Document type
Journal article
Publisher
IOP Publishing
Topic
- Atom and Molecular Physics and Optics
Status
Published
ISBN/ISSN/Other
- ISSN: 0957-0233