Measurement of Tumor Extent and Effects of Breast Compression in Digital Mammography and Breast Tomosynthesis
Author
Summary, in English
Tumor size is an important predictor of prognosis and treatment effect. We hypothesized that the tumor outline would be better defined in BT and therefore tumor measurement in BT would be more accurate compared with DM. The results showed that breast tumor size measured on BT correlated better with the size measured by the pathologists on the surgical specimens compared with measurement on DM.
Breast compression is important in mammography both to improve image quality and to reduce the radiation dose to the breast, but it also has a negative consequence as some women refrain from mammography due to the pain associated with the examination. Since BT is a 3D technique, it was hypothesized that less breast compression force can be applied. The results indicated that less compression force is possible without significantly compromising the diagnostic quality of the image and that the patient comfort was improved.
An applied breast compression force as used in mammography results in a pressure distribution over the breast. The pressure distribution was assessed using thin pressure sensors attached to the compression plate. The results showed that the pressure distribution was heterogeneous in appearance and varied widely between different breasts. In almost half of the subjects most of the pressure was over the juxtathoracic part of the breast and the pectoral muscle with little or no pressure over the rest of the breast.
Another concern regarding breast compression is the question whether the resulting pressure might damage tumors, causing a shedding of malignant cells into the blood system. Peripheral venous blood samples were drawn before and after breast compression and analyzed for circulating tumor cells. The study found no elevated number of circulating cancer cells in peripheral blood after breast compression. Future analysis of samples from veins draining the breast are needed to study if circulating tumor cells are being trapped in the lung capillaries.
Department/s
Publishing year
2013
Language
English
Publication/Series
Lund University Faculty of Medicine Doctoral Dissertation Series
Volume
2013:58
Full text
- Available as PDF - 13 MB
- Download statistics
Document type
Dissertation
Publisher
Medical Radiation Physics, Lund University
Topic
- Radiology, Nuclear Medicine and Medical Imaging
Keywords
- digital mammography
- breast tomosynthesis
- breast compression
- pressure distribution
- breast cancer
- circulating tumor cell
Status
Published
Research group
- Medical Radiation Physics, Malmö
ISBN/ISSN/Other
- ISSN: 1652-8220
- ISBN: 978-91-87449-28-4
Defence date
17 May 2013
Defence time
13:00
Defence place
Room 2005-2007, Inga Marie Nilssons gata 49, SUS Malmö
Opponent
- Andrew Maidment