Polymeric integrated selective enrichment target (ISET) for solid-phase-based sample preparation in MALDI-TOF MS
Author
Summary, in English
d A polymer microfabricated proteomic sample preparation and MALDI MS sample presentation device, the integrated selective enrichment target (ISET), comprising an array of perforated nanovials is reported. Each perforated nanovial can be filled with selective extraction media (microbeads) for purification and concentration of protein/peptides prior to matrix-assisted laser desorption/ionization mass spectrometry (MALDI MS). The main areas covered are the influence of the molding-process-induced surface roughness and how to address the lack of inherent conductivity in the polyetheretherketone (PEEK) material for optimal MALDI MS readout. Application of the disposable polymeric ISET devices for solid-phase extraction and phosphopeptide capture is also demonstrated.
Publishing year
2007
Language
English
Pages
1445-1452
Publication/Series
Journal of Mass Spectrometry
Volume
42
Issue
11
Document type
Journal article
Publisher
John Wiley & Sons Inc.
Topic
- Medical Engineering
- Analytical Chemistry
Keywords
- polymeric MALDI
- microfluidics
- SPE
- MALDI MS
- solid-phase extraction
- ISET
- target
Status
Published
ISBN/ISSN/Other
- ISSN: 1076-5174