A flexible wavelet filter structure for cardiac pacemakers: a power efficient implementation

Author

Summary, in English

This paper describes the modification of a wavelet based event detector for cardiac pacemakers with improved power efficiency. A previously published R-wave detector has been supplemented with a noise detector that operates in supervision mode, and automatically activates and deactivates parts of the detector. Simulation results show that the power consumption in an ASIC core can be reduced by more than 65% if the patient is at rest or in a low-noise environment, whereas reliable detection performance is sustained at moderate to high noise levels. The detector has been synthesized in 0.13 μm CMOS technology

Publishing year

2004

Language

English

Pages

1-8

Publication/Series

[Host publication title missing]

Document type

Conference paper

Publisher

IEEE - Institute of Electrical and Electronics Engineers Inc.

Topic

  • Electrical Engineering, Electronic Engineering, Information Engineering

Keywords

  • flexible wavelet filter structure
  • event detector
  • R-wave detector
  • noise detector
  • ASIC
  • improved power efficiency
  • CMOS technology
  • cardiac pacemakers

Conference name

IEEE International Workshop on Biomedical Circuits and Systems (BioCAS)

Conference date

2004-12-01 - 2004-12-03

Conference place

Singapore

Status

Published

ISBN/ISSN/Other

  • ISBN: 0-7803-8665-5