Classification of Electrocardiographic P-wave Morphology

Author

Summary, in English

The atrial activity of the human heart is normally visible in the electrocardiogram as a P-wave. In patients with intermittent atrial fibrillation, a different P-wave morphology can sometimes be seen, indicating atrial conduction defects. The purpose of this study was to develop a method to discriminate between such P-waves and normal ones. 20 recordings of each type were used in a classification which, based on impulse response analysis of the P-wave and linear discrimination between various parameters, produced a correct classification in 37 of the 40 recordings (sensitivity 95%, specificity 90%)

Publishing year

2001

Language

English

Pages

401-405

Publication/Series

IEEE Transactions on Biomedical Engineering

Volume

48

Issue

4

Document type

Journal article

Publisher

IEEE - Institute of Electrical and Electronics Engineers Inc.

Topic

  • Medical Engineering

Keywords

  • medical signal processing
  • electrocardiography
  • identification

Status

Published

ISBN/ISSN/Other

  • ISSN: 1558-2531