High-resolution spectroscopy of laser-produced plasmas in the photon energy range above 10 keV.

Author

Summary, in English

X-rays from a laser-produced plasma were analysed using single-photon counting germanium detectors. Special attention was given to the extremely high photon flux, requiring either a long source-detector distance and narrow collimated measurement geometry in order to avoid severe pile-up of pulses in the spectrometer, or the use of a Compton scattering geometry. With the use of these techniques, individual characteristic K-lines were resolved for the first time. Efficiencies for the conversion of the laser light into X-rays were determined, being up to 10−4 for Bremsstrahlung, and up to 10−5 for the characteristic K-line emission. Electron temperatures of not, vert, similar100 keV were assessed.

Publishing year

1997

Language

English

Pages

387-396

Publication/Series

Nuclear Instruments & Methods in Physics Research. Section A: Accelerators, Spectrometers, Detectors, and Associated Equipment

Volume

394

Issue

3

Document type

Journal article

Publisher

Elsevier

Topic

  • Accelerator Physics and Instrumentation

Status

Published

Research group

  • Medical Radiation Physics, Malmö

ISBN/ISSN/Other

  • ISSN: 0167-5087