The browser you are using is not supported by this website. All versions of Internet Explorer are no longer supported, either by us or Microsoft (read more here: https://www.microsoft.com/en-us/microsoft-365/windows/end-of-ie-support).

Please use a modern browser to fully experience our website, such as the newest versions of Edge, Chrome, Firefox or Safari etc.

First certified reference material for cystatin C in human serum ERM-DA471/IFCC.

Author

Summary, in English

The IFCC Working Group for the Standardisation of Cystatin C (WG-SCC), in collaboration with the Institute for Reference Materials and Measurements (IRMM), announces the availability of the new certified reference material ERM-DA471/IFCC. The material was characterised using a pure protein primary reference preparation (PRP) as calibrant. The PRP was prepared from recombinant cystatin C, and its concentration measured using dry mass determination. The characterisation of ERM-DA471/IFCC was performed by particle enhanced immuno-nephelometry, particle enhanced immuno-turbidimetry, and enzyme amplified single radial immuno-diffusion. The certified cystatin C mass concentration in ERM-DA471/IFCC, if reconstituted according to the specified procedure, is 5.48 mg/L, the expanded uncertainty (k=2) being 0.15 mg/L.

Publishing year

2010

Language

English

Pages

1619-1621

Publication/Series

Clinical Chemistry and Laboratory Medicine

Volume

48

Issue

11

Document type

Journal article

Publisher

De Gruyter

Topic

  • Clinical Laboratory Medicine

Status

Published

ISBN/ISSN/Other

  • ISSN: 1434-6621