A comparison between one first generation and three second generation anti-HCV ELISAs: an investigation in high- and low-risk subjects in correlation with recombinant immunoblot assay and polymerase chain reaction

Author

Summary, in English

One first generation assay (manufactured by Ortho, test I) and 3 second generation anti-HCV ELISAs (manufactured by Ortho, Abbott, and UBI, tests II-IV) were compared. Sera from 4 different sources were used: (1) intravenous drug-users (IVDUs, n = 50), (2) blood donors (n = 1055), (3) all clinical samples from one day of routine anti-HCV testing (n = 89), (4) hemodialysis patients previously found negative by test I but clinically suspected to have a HCV infection (n = 11). Confirmatory anti-HCV tests were carried out with a second generation recombinant immunoblot assay (RIBA II). In sera positive exclusively by test IV, one antibody consumption test (UBI HCV Neutralization EIA) and one further immunoblot assay (INNO-LIA HCV Ab) were used. PCR for HCV RNA was carried out on all hemodialysis patient sera and in the RIBA II positive blood donor sera. The second generation ELISAs discriminated 11 more positive samples than the first generation test (2 IVDUs, 5 blood donors, 4 clinical samples). The 9 sera from blood donors and clinical samples were all RIBA II positive or indeterminate. The second generation tests thus showed increased sensitivity. The second generation tests also showed increased specificity in that 4 samples that were positive by test I but negative by the second generation tests, were also negative by RIBA II. With few exceptions, all RIBA II-positive and most of the indeterminate samples were positive by the second generation ELISAs. With few exceptions, all the RIBA II-negative samples were negative by the second generation ELISAs. Eleven blood donor sera were positive by test IV exclusively where RIBA II and other supplementary assays were negative. The recently introduced second generation anti-HCV ELISAs were found to have a higher sensitivity than the first generation test. The tests also showed a good concordance with the exception of test IV in the group of blood donor sera.

Publishing year

1992

Language

English

Pages

287-296

Publication/Series

Journal of Virological Methods

Volume

40

Issue

3

Document type

Journal article

Publisher

Elsevier

Topic

  • Microbiology in the medical area

Keywords

  • Anti-HCV
  • ELISA
  • Recombinant immunoblot assay
  • Polymerase chain reaction

Status

Published

Research group

  • Clinical Microbiology, Malmö

ISBN/ISSN/Other

  • ISSN: 1879-0984