Structural Parameters determined by microCT in normal and osteoporotic human trabecular Bone are Affected Differently by image resolution

Author

  • Hanna Isaksson
  • Juha Töyräs
  • Mikko Hakulinen
  • Inari Tamminen
  • Antti Aula
  • Petro Julkunen
  • Heikki Kröger
  • Jukka Jurvelin

Summary, in Swedish

Abstract in Undetermined

This study employed microCT to investigate whether image resolution affects bone structural parameters differently in healthy normal and osteoporotic trabecular bone. With increasing image voxel size, the originally detected differences between sample groups diminished. The results suggest that structural differences may not be reliably detected with clinical scanners.

Publishing year

2011

Language

English

Pages

167-177

Publication/Series

Osteoporosis International

Volume

22

Issue

1

Document type

Journal article

Publisher

Springer

Topic

  • Orthopedics

Status

Published

ISBN/ISSN/Other

  • ISSN: 1433-2965